Resulting Calibration Test Points

As discussed in Test Point Grid Setup, the final structure of the test points is different from the setup grid.

In this example, you have several options for picking 100 to 600 microliters on a 5-channel pipette (configured in Pipette Method Setup).

Test point grid setup options

Test point grid setup options

After running the Minimal group, you get this result:

Resulting calibration test points

Resulting calibration test points

A copy of each setup test point is generated four times — one each for mean, standard deviation, accuracy, and precision — and then once per channel. For any given test point, dragging the right-hand slider over reveals the individual As Found and As Left samples that generated the data.

To understand how this is collapsed into a more compact form on the certs, see Explanation of Pipette Certs.