Skip to main content

Test Point Grid Setup

The first two steps in setting up a pipette's test points are:

  1. Set the Setup Grid and Results Grid to PIPETTE.
  2. Choose one of the Pipette options as the Tolerance Type.

Because pipettes are defined by a nominal value with an accuracy and/or precision, the test-points grid looks a little non-standard.

Pipette test-points grid

Pipette test-points grid

Tolerance types

TypeInterpretation
PipetteLegacy from 12.1 and earlier. Treated as Pipette % Value.
Pipette % RangeAccuracy and Precision are interpreted as a percentage of the maximum nominal value for that group.
Pipette % ValueAccuracy and Precision are interpreted as a percentage of the current test point.
Pipette microLAccuracy and Precision are interpreted as the limit in microliters.

Test-point grid columns

ColumnDescription
Group (Custom 1)Lets you have several predefined sets of test points to choose from at calibration time.
Af Channels (Custom 2)A subset of the channels to test for As Found, by default. Should be the same for all items in a given group. Blank defaults to all channels. 0 indicates As Found data is not collected for this test point in this group.
Al Channels (Custom 3)Same as Af Channels but for As Left.
Af Samples (Custom 4)A subset of the samples to test for As Found, by default. Should be the same for all items in a given group. Blank defaults to the value selected in Pipette Method Setup.
Al Samples (Custom 5)Same as Af Samples but for As Left.
DescriptionA good description of the line.
StandardAlways the target in microliters.
Accuracy (Extra Num 1)Desired percent accuracy (between 0 and 100%).
Alt. Accuracy (Extra Num 3)Optional. Reminds the technician of an alternate accuracy. For example, if you calibrate to manufacturer spec with ISO 8655 as a backup, list the ISO 8655 spec here.
Precision (Extra Num 2)Desired percent precision (between 0 and 100%).
Alt. Precision (Extra Num 4)Optional. Same idea as Alt. Accuracy for precision.
Res.Desired resolution.
Uncertainty Res.Desired uncertainty resolution.
UnitAll pipette calculations use microliters internally. Putting a different unit here generally causes confusion.

Per-group channel selection

Per-group channel selection example

Per-group channel selection example

Take care if you want only certain As Found / As Left test points in a group. In the screenshot above:

  • Group af1256 uses test points 1–2 and 5–6 for As Found, and 3–4 for As Left.
  • Group af136 uses 1, 3, 6 for As Found, and 2, 4–5 for As Left.

Use a Pipette Tolerance Type to ensure dual-tolerancing works properly.

Resulting test points

When the pipette data is complete, the results grid will look like this:

Resulting test points grid

Resulting test points grid

Each default test point generates 4 test points per channel. So a 4-channel pipette with 5 setup test points produces 4 × 5 × 4 = 80 test points in the resulting calibration.

As Found / As Left readings are stored as either 1 (in tolerance) or 0 (out of tolerance) for the purposes of determining the overall calibration result.

Saved environmental data

As Found and As Left environmental data is saved in the following fields. (Replace AF with AL for As Left.)

FieldStores
ENV_AF_BPBarometric Pressure
ENV_AF_BP_TYPEBarometric Pressure Units
ENV_AF_RHRelative Humidity
ENV_AF_TEMPTemperature
ENV_AF_TEMP_TYPETemperature Units
AF_EXTRANUM1Z * Y Factor
AF_EXTRANUM2Air density
AF_EXTRANUM3Cubic Expansion
AF_EXTRANUM4Y-Factor